214 Endress+Hauser
719 (A719) Alarm
B
B>Y-Wert
(LINEARIZATION_TAB
LE_Y_VALUE) of lin.
table out of edit limits
19 7 At least one Y-Wert
(LINEARIZATION_TABLE_Y_
VALUE) in the linearization
table is below the Scale Out/
SCALE_OUT, EU at 0%/EU_0,
TANK CONTENT MIN. or
above the Scale Out/
SCALE_OUT, EU at 100%/
EU_100, TANK CONTENT
MAX.
Carry out calibration again.
(See also these Operating
Instructions, Section 4 and
Section 8)
40
720 (E720) Error
C
C>Sensor under
temperature
17 0 The temperature measured in
the sensor is smaller than the
lower nominal temperature of
the sensor. ( ä 76 and
ä 199, parameter
description for Tmin Sensor/
SENSOR_TEMP_LOW_LIMIT)
Increase process temperature/
ambient temperature.
34
Unsuitable download carried
out.
Check configuration and
perform download again.
721 (A721) Alarm
B
B>Zero Position
(ZERO_POSITION) level
out of edit limits
18 7 Level Min/LEVEL_MIN or
Level Max/LEVEL_MAX has
been changed.
Perform reset (code 2710) and
recalibrate the device.
41
722 (A722) Alarm
B
B>EMPTY CALIB. or
FULL CALIB. are out of
edit limits
18 7 Level Min/LEVEL_MIN or
Level Max/LEVEL_MAX has
been changed.
Perform reset (code 2710) and
recalibrate the device.
42
723 (A723) Alarm
B
B>Max. Flow/
FLOW_MAX
(Scale Out/SCALE, EU
at 100%/EU_100) out of
edit limits
18 7 Flow-Meas. Type/FLOW_
MEAS_TYPE has been changed.
Carry out calibration again. 43
725 (A725) Alarm
B
B>Sensor connection
error, cycle disturbance
20 5 Electromagnetic effects are
greater than specifications in the
technical data.
See Technical Information
TI00382P (Deltabar S),
TI00383P (Cerabar S) or
TI00416P (Deltapilot S).
Setscrew loose.
Block off electromagnetic effects
or eliminate source of
disturbance.
Retighten setscrew with 1 Nm
(0,74 lbf ft) (see chapter
"Rotating the housing" in
BA00301P (Deltabar S) or
BA00302P (Cerabar S) or
BA00372P (Deltapilot S)).
26
Sensor or main electronics
defect.
Replace sensor or main
electronics.
726 (E726) Error
C
C>Sensor temperature
error - overrange
20 5 Electromagnetic effects are
greater than specifications in the
technical data.
See Technical Information
TI00382P (Deltabar S),
TI00383P (Cerabar S) or
TI00416P (Deltapilot S).
Block off electromagnetic effects
or eliminate source of
disturbance.
32
Process temperature is outside
permitted range.
Check temperature present,
reduce or increase if necessary.
Sensor defect. If the process temperature is
within the permitted range,
replace sensor.
Code Error
type/
NA 64
Message/description XD_
ERROR
Value
BLOCK_
ERROR
Bit
Cause Measure Prio-
rity