Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Troubleshooting
Endress+Hauser 217
746 (W746) Warning
C
C>Sensor connection
error - initializing
20 5 Electromagnetic effects are
greater than specifications in the
technical data.
See Technical Information
TI00382P (Deltapilot S),
TI00383P (Cerabar S) or
TI00416P (Deltapilot S). This
message normally only appears
briefly.
Wait a few minutes.
Restart the device. Perform reset
(Code 7864).
Block off electromagnetic effects
or eliminate source of
disturbance.
27
Overpressure or low pressure
present.
Reduce or increase pressure.
747 (A747) Alarm
B
B>Sensor software not
compatible to electronics
17 0 Sensor does not suit the device
(electronic sensor nameplate).
Replace sensor with a suitable
sensor.
17
748 (A748) Alarm
B
B>Memory failure in
signal processor
20 5 Electromagnetic effects are
greater than specifications in the
technical data.
See Technical Information
TI00382P (Deltapilot S),
TI00383P (Cerabar S) or
TI00416P (Deltapilot S).
Block off electromagnetic effects
or eliminate source of
disturbance.
16
Main electronics defect. Replace main electronics.
750 (A750) Alarm
B
B>Configuration not
permitted
19 7 By means of operation, options
were selected for the
configuration of the device that
do not suit one another. e.g. if
the "1 (linearization table)"
option was selected for
Linearization/LINEARIZATION
and the "1347 (m
3
/s)" unit for
Primary Value Type/PRIMARY_
VALUE_UNIT.
Check configuration.
Perform reset (code 7864) and
recalibrate the device.
45
Code Error
type/
NA 64
Message/description XD_
ERROR
Value
BLOCK_
ERROR
Bit
Cause Measure Prio-
rity