Parameter description (FF configuration program) Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus
82 Endress+Hauser
7.3.4 DP Flow Transducer Block (only Deltabar S)
DP Flow Transducer Block
Parameter Description
Device Dialog/
DEVICE_DIALOG
Display
Index: 11
Data type: unsigned8
Access: read only
If the configuration is not suitable, this parameter displays a message saying that a
configuration error is present. The message can indicate which parameter was incorrectly
configured.
Insert Pin No./
SW_LOCK
Entry
Index: 12
Data type: unsigned16
Access: read, write
For entering a code to lock or unlock operation.
!
Hinweis!
• The symbol on the on-site display indicates that operation is locked. Parameters which
refer to how the display appears, e.g. LANGUAGE and DISPLAY_CONTRAST can still be
altered.
• If operation is locked by means of the DIP-switch, cou can only unlock operation again by
means of the DIP-switch. If operation is locked by means of remote operation, you can
only unlock operation again by means of remote operation.
→ See also Operating Instructions BA00301P (Deltabar S) or BA00302P (Cerabar S) oder
BA00372P (Deltapilot S) Chapter "Locking/unlocking operation".
Options:
• Lock: enter a number between 0...9999 which is ≠100.
• Unlock: enter number 100.
Factory setting:
100
Status Locking/
STATUS_LOCKING
Index: 13
Data type: unsigned8
Access: read only
Shows the actual locking status of the device or conditions that can cause the locking.
(Hardware locking, Software locking)
Flow-Meas. Type/
FLOW_MEAS_TYPE
Selection
Index: 14
Data type: unsigned8
Access: read, write, OOS
Select the flow type.
Prerequisite:
• Deltabar S differential pressure transmitter
Options:
• Volume operat. cond. (volume under operating conditions)
• Volume norm. cond. (norm volume under norm conditions in Europe: 1013.25 mbar and
273.15 K (0 °C))
• Volume std. cond. (standard volume under standard conditions in USA: 1013.25 mbar
(14.7 psi) and 288.15 K (15 °C/59 °F))
• Mass p. cond. (mass under operating conditions)
Factory setting:
Volume operat. cond.
Suppressed Flow/
SUPPRESSED_FLOW
Display
Index: 15
Data type: Float
Access: read only
Displays the current flow. Depending on the flow mode selected (
→ Flow-Meas. Type/
FLOW_MEAS_TYPE), a volume flow, mass flow, standard volume flow or corrected volume
flow is
displayed.