Mounting Levelflex M FMP40 FOUNDATION Fieldbus
30 Endress+Hauser
Installing FMP40 with heat insulation
If process temperatures are high, FMP40 must be included in normal tank insulation to prevent
the electronics heating up as a result of heat radiation or convection.
The insulation may not exceed beyond the points labelled "MAX" in the drawing.
Process connection with adapter
G¾, G1½, ¾NPT or 1½NPT
Process connection with flange
DN40 to DN200
L00-FMP40xxx-17-00-00-en-003 L00-FMP40xxx-17-00-00-en-002
MAX
max. 150° C (302° F)
tank insulation
MAX
40
max. 150° C (302° F)
tank insulation
Installation at an angle
For mechanical reasons, the probe should be
installed as vertically as possible.
With inclined installations the probe length
has to be adjusted in dependence to the instal-
lation angle.
–up to 1m = 30°
–up to 2m = 10°
–up to 4m = 5°
L00-FMP4xIxx-17-00-00-de-048