Levelflex M 8 Function group "output" (06), - "PROFIBUS param." (06), PROFIBUS PA only
Endress+Hauser 53
8.11 Function "simulation" (065)
If necessary, linearisation, the output signal and the current output can be tested with the simulation
function. You have the following simulation options:
Selection:
• sim. off
• sim. level
• sim. volume
• sim. current (HART only)
sim. off
Simulation is switched off.
sim. level
Enter the level value in "simulation value" (066).
The functions
• measured value (000)
• measured level (0A6)
• output current" (067) - only with HART instruments!
follow the entered values.
sim. volume
Enter the volume value in "simulation value" (066).
The functions
• measured value (000)
• output current" (067) - only with HART instruments!
follow the entered values.
sim. current (HART only)
Enter the current value in "simulation value" (066).
The function
• output current" (067) - only with HART instruments!
follows the entered values.
⇒