15 Trouble-shooting Levelflex M
78 Endress+Hauser
15.3 Application errors
Error Output Possible cause Remedy
A warning or alarm
has occurred.
Depending on the configuration See table of
error messages
(
ä 76)
1. See table of error messages
(
ä 76)
Measured value
(00) is incorrect
L00-FMP4xxxx-19-00-00-en-019
Measured distance
(008) OK?
yes
1. Check empty calibr. (005) and
full calibr. (006)
2. Check linearization:
Level/ullage (040)
Max. scale (046)
Diameter vessel (047)
Check table
no
An interference echo
may have been
evaluated.
yes
1. Carry out tank mapping
Basic setup
No change of
measured value on
filling/emptying
L00-FMR2xxxx-19-00-00-en-014
Interference echo from
tank internals, nozzle
or buildup on the
probe
1. Carry out tank mapping
Basic setup
2. If necessary, clean probe
3. If necessary, select better
mounting position
E 641 (loss of
echo) after
switching on the
power supply
If the device is configured to HOLD in the event
of echo loss the output is set to any
value/current.
Noise level during the
initialization phase too
high.
Repeat empty calibr. (005) once
more.
Caution!
Before confirmation change with
O
or
S
to the edit mode.
20 mA/100%
F m/ft
E m/ft
4 mA/0% t
D m/ft (008)
actual
expected
20 mA/100%
4 mA/0% t
actual
expected