Operation Liquistation CSF34
114 Endress+Hauser
Menu/Setup/Inputs
Function Options Info
Interval Selection
• Daily
• Weekly
• Monthly
Factory setting
Daily
Daily
If a daily interval is selected, set the Time in
the following menu item.
Weekly
If a weekly interval is selected, set the Day
of week and the Time in the following
menu items.
Monthly
If a monthly interval is selected, set the Day
of month and the Time in the following
menu items.
Time 00:00:00 ... 23:59:59
HH:MM:SS
Factory setting
12:00:00 HH:MM:SS
Having selected Input variable Rainfall
Signal slope Selection
• Low-High
• High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Unit Selection
• mm
• inch
Factory setting
mm
Select the unit.
Meas. value format Factory setting
#.#
Specify the number of decimal places.
1 pulse = 0.00 to 5.00 mm
Factory setting
1.0 mm
Definition of the pulse value, limits are
calculated depending on the unit. The
correct switch value is provided in the
Operating Instructions of your rain gage.
Intensity Selection
• mm/min
• mm/h
• mm/d
Factory setting
mm/min
Select the intensity per minute, hour or day
according to your requirements.
Totalized rainfall
Totalized rainfall - - - The totalized rainfall is displayed.
Reset totalizer Selection
• Manual
• Automatic
• At program start
Factory setting
Manual
Manual
Reset the counter manually.
Automatic
The counter is reset automatically at
intervals.
At program start
The counter is reset at program start.
Having selected Reset totalizer Manual
Reset totalized rainfall Action The totalized rainfall currently calculated is
set to zero when the meter is reset
manually.
Having selected Reset totalizer Automatic