Operation Liquistation CSF34
124 Endress+Hauser
Function = Diagnostics
Function Options Info
Operating mode Selection
• as assigned
• Namur M
• Namur S
• Namur C
• Namur F
Factory setting
• Relay:
as assigned
• Alarm relays:
Namur F
as assigned
If this option is selected, the diagnostic messages
which you have individually assigned to the relay
are output via the relay.
Namur M ... Namur F
If you decided to use one of the Namur classes,
all the messages that are assigned to the
individual class are output via the relay. You can
also change the Namur class assignment for
every diagnostic message.
(Menu/Setup/General settings/Extended
setup/Diagnostics settings/Diag. behavior or
Menu/Setup/Inputs/<Sensor>/Extended
setup/Diagnostics settings/Diag. behavior)
Attributed diagnostic messages
Operating mode = as
assigned
Read only All the messages assigned to the relay output
appear on the display. You do not have the
option of editing the information here.
Outputting the status of a cleaning function
Function = Cleaning
Function Options Info
Assignments Selection
• None
• Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
You have the following options depending on the
cleaning program that is selected (Menu/Setup/
Additional functions/Cleaning:
• Cleaning type = Standard clean
Cleaning 1 - Water, Cleaning 2 - Water,
Cleaning 3 - Water, Cleaning 4 - Water
• Cleaning type = Chemoclean
Cleaning 1 - Water, Cleaning 1 - Cleaner,
Cleaning 2 - Water, Cleaning 2 - Cleaner,
Cleaning 3 - Water, Cleaning 3 - Cleaner,
Cleaning 4 - Water, Cleaning 4 - Cleaner
• Cleaning type = Chemoclean Plus
4x Cleaning 1 - %0V, 4x Cleaning 2 - %0V
1)
Hold behavior Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Ignore
Freeze last value
The device freezes the last measured value.
Fixed value
You define a fixed measured value that is output
at the output.
Ignore
A hold has no effect.
1) %0V is variable text which you can assign in Menu/Setup/Additional functions/Cleaning/Chemoclean
Plus/Output label 1 ... 4.
10.5.4 HART
Specify which device variables should be output via HART communication.
You can define a maximum of 16 device variables.
1. Specify the data source.
You can choose from sensor inputs and controllers.
2. Select the measured value to be output.