Diagnostics and troubleshooting Liquistation CSF34
164 Endress+Hauser
Entry Example Info
Running sample number 1 Running sample number in the current
program
Flow sum since last sampling 1 For flow-paced and time/flow-paced
sampling:
--> Flow since the last sampling
For all other types of sampling:
--> Display: 0
11.9.4 Bottle statistics
Display shows the bottle statistics of the sampler:
‣
In Menu/Diagnostics/Logbooks/Logbook program select the Show summary of
current program item or select the MEAS soft key; works both when the program is
active and when it has stopped.
The statistics appears for each individual bottle when the program is started. This
gives you detailed feedback on the last sampling operations.
The statistics are deleted when the following event occurs:
Program start
The statistics are selectively overwritten when the following event occurs:
When the first bottle is reached in situations where "Continuous operation" is
configured as the end of the program in the program settings
The statistics are displayed as follows:
A0045690
Display Info
bt The bottle number is displayed.
hh:mm The time the first sample was transferred to the bottle is displayed.
DD-hh:mm The time the first sample was transferred to the bottle is displayed.
Smp Displays how often sampling was triggered per bottle.
n.s Indicates the number of times a sample was not taken even though sampling was
triggered. This can occur if the maximum permissible fill volume for the bottle has
been reached but the system is still supposed to transfer samples to the bottle. The
"Overfill sensor" message is displayed while the program is active.
n.f The value indicates how often sampling was canceled because it was not possible to
draw in any or enough medium into the dosing glass to cover the LF1 probe.
ml The sampling volume collected per bottle is displayed.
Q The total flow for every bottle is displayed (only if flow measurement is connected).