Liquistation CSF34 Diagnostics and troubleshooting
Endress+Hauser 163
Entry Example Info
Sampling mode Time-paced CTCV Time-paced CTCV
Time-paced
Flow-paced VTCV
Flow-paced
Time/flow-paced CTVV
Time/flow-paced
Single sample
Single sample
Sample table
Single sample
Sampling mode
Display of the sampling mode
Sampling interval/unit 10 min Display of the interval and the unit
Samples/bottle 4 With bottle change
Number of samples per bottle
.
Bottles/sample 0 Multiple bottles,
Sampling volume/unit 100 ml Sample volume when sampling
Start mode Immediate Field only populated for PrgPartStart,
BasicPrgStart and StdPrgStart:
--> The program start setting is displayed
• Immediate --> immediately
• Date/time --> after date/time
• Volume --> with a volume
• Event --> when an event occurs
• Interval --> after an interval
• Individual dates --> individual timetable
• Multiple date --> multiple dates
Start date 05.05.2010 Field only populated if Start mode = Date/
Time:
--> The start date is displayed
Stop mode Program end The program stop setting is displayed:
• Program end --> when the program ends
• Continuous --> continuous operation
• Bottles full --> when bottles are full
• Date/time --> after date/time
• Event --> when an event occurs
Stop date 06.05.2010 Field only populated if Program end =
Date/Time:
--> The date the program was stopped is
displayed
Start flow sum/unit 100 m
3
Field only populated if Start mode =
Volume:
--> The starting volume is displayed
Bottle number 1 Field only populated for BasicSampling or
StdSampling:
--> The bottle which was filled with the
sample is displayed
Sample nbr 2 Number of samples transferred to the
current bottle
Sampling result Sampling Ok Sampling Ok --> sampling ok
Sampling nOk --> sampling failed
--> For detailed diagnostics messages, see
the diagnostics logbook