Diagnostics and troubleshooting Liquistation CSF34
162 Endress+Hauser
DIAG/Logbooks
Function Options Info
Delete all entries Action You can use this to delete all data logbook
entries.
Save logbooks
File format Selection
CSV
FDM
Save the logbook in the preferred file
format.
You can then open the saved CSV file on the PC
in MS Excel, for example, and make further edits
here.
1)
. You can import the FDM files into
FieldCare and archive them so that they are
tamper-proof.
Program logbook
All data logbooks
Data logbook 1 ... 8
All event logbooks
Calibration logbook
Diagnostic logbook
Configuration logbook
HW version logbook
Version logbook
Action,
commences as soon as the
option is selected
Use this function to save the logbook to an SD
card.
Insert the SD card into the device card reader
and select the logbook to be saved.
Save the logbook in the preferred file
format. You can then open the saved CSV
file on the PC in MS-Excel, for example, and
edit it. You can import the FDM files into
Fieldcare and archive them so they are
tamper-proof.
The file name is made up of the Logbook ident (Menu/Setup/General settings/Logbooks), an
abbreviation for the particular logbook and a time stamp.
1) CSV files use international number formats and separators. Therefore they must be imported into MS
Excel as external data with the correct format settings. If you double-click the file to open it, the data are
only displayed correctly if MS Excel is installed with the US country setting.
11.9.3 Program logbook
Entry Example Info
Timestamp 05.05.2010 12:40 Time stamp - the start time in the case of
sampling
Event BasicPrgStart Power on
--> Time the device is started
Power failure
--> Time the power failed (to the minute)
BasicPrgStart, StdPrgStart
--> Time the program was started
BasicSampling, StdSampling
--> Entry made during sampling
PrgPartStart, PrgPartStop
--> Time a subprogram is enabled and
disabled
PrgStop
--> Time the program was ended
Name Program1 In the case of BasicPrgStart, StdPrgStart,
BasicSampling or PrgStop
--> The name of the program appears
In the case of StdSampling, PrgPartStart
or PrgPartStop
--> The name of the subprogram appears
Bottle configuration 12x+6x - PE/glass plate
distribution
The selected bottle configuration is
displayed
Left bottle volume 1000 The bottle volume is displayed
--> "Right bottle volume" remains empty for
bottle configurations with different volumes
Right bottle volume 3000