EasyManua.ls Logo

Endress+Hauser Liquistation CSF34 - Wiring Cables at Terminals

Endress+Hauser Liquistation CSF34
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Liquistation CSF34 Electrical connection
Endress+Hauser 33
6.4.2 Wiring cables at cable terminals
Plug-in terminals for Memosens and PROFIBUS/RS485 connections
Press the screwdriver against
the clip (opens the terminal).
Insert the cable until the limit
stop.
Remove the screwdriver (closes
the terminal).
After connection, make sure that every cable end is securely in place. Terminated
cable ends, in particular, tend to come loose easily if they have not been correctly
inserted as far as the limit stop.
All other plug-in terminals
Press the screwdriver against
the clip (opens the terminal).
Insert the cable until the limit
stop.
Remove the screwdriver (closes
the terminal).
After connection, make sure that every cable end is securely in place. Terminated
cable ends, in particular, tend to come loose easily if they have not been correctly
inserted as far as the limit stop.
 35 Press the screwdriver
against the clip (opens the
terminal)
 36 Insert the cable until the
limit stop
 37 Remove the screwdriver
(closes the terminal)