Electrical connection Liquistation CSF34
34 Endress+Hauser
6.4.3 Module 485DP
A0050399
38 Module
81
82
DIP switch
on
1
T
COM
SF
BF
PWR
DGND
VP
96'
95'
99'
Termi-
nation
DP
96
95
99
2
4
8
16
32
64
128/SW
Service
A0050400
39 Wiring
diagram
Terminal PROFIBUS DP
95 A
96 B
99 Not connected
82 DGND
81 VP
LEDs on front of module
LED Designation Color Description
PWR Power GN Supply voltage is applied and module is initialized.
BF Bus failure RD Bus failure
SF System failure RD Device error
COM Communication YE PROFIBUS message sent or received.
T Bus termination YE • Off = No termination
• On = Termination is used
DIP switches on front of module
DIP Factory setting Assignment
1-128 ON Bus address (→ "Commissioning/communication")
OFF Write protection: "ON" = configuration not possible via the bus, only via local
operation
Service OFF The switch has no function