Liquistation CSF34 Commissioning
Endress+Hauser 57
9.3.5 Sampling programs
Difference between program types
The following box provides an overview of the differences between the Basic, Standard and
Advanced program types.
Basic (1 sampling program)
Start condition:
Immediate
Date/time
Immediate activation
Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
external signal,
Bottle change after time or
number of samples, external
signal
Bottle synchronization
Multiple bottles
Stop condition:
Program end
Continuous operation
Standard (1 sampling program with 1-5 sub-programs)
Start condition:
Immediate
Date/time
Volume
Immediate activation, individual
times, multiple times, interval,
deactivation of sub-program 1
Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
external signal
Bottle change after time or
number of samples, external
signal
Bottle synchronization
Multiple bottles
Stop condition:
Program end
Continuous operation
Date/time
Advanced (1 sampling program with 1-24 sub-programs)
Start condition:
Immediate
Date/time
Volume
External signal
Fieldbus
Immediate activation, individual
times, multiple times, interval,
event, external start,
deactivation of sub-program 1,
fieldbus
Time-paced, volume-paced or
flow-paced (CTCV, VTCV, CTVV),
single sample, sample table,
external signal, fieldbus
Bottle change after time or
number of samples, external
signal, fieldbus
Sample synchronization
Bottle synchronization
Multiple bottles
Stop condition:
Program end
Continuous operation
Date/time
Fieldbus