Commissioning Liquistation CSF34
58 Endress+Hauser
Manual sampling
1. Trigger manual sampling with the MAN soft key. This pauses any program currently
running.
The current bottle configuration and the current sample volume are displayed.
You can select the distributor position. In the case of peristaltic systems, you can
also change the sample volume.
In the case of vacuum systems, a multiple of a single manual sample can be taken
under Multiplier. Specify the Multiplier range from 1 to 50.
2. Select Start sampling.
A new screen is displayed indicating the progress of the sampling process.
3. After a manual sampling has been performed, press ESC to display and resume an
active program.
The sample volume for "Manual sampling" is not taken into account in the
calculated bottle volumes.
Programming for automatic sampling
In the overview screen, create a simple sampling program under Select sampling
program/New/Basic or under Menu/Setup/Sampling programs/Setup program/New/
Basic:
1. Enter the "Program name".
2. The settings from the Basic setup for the bottle configuration and bottle volume are
displayed.
3. Sampling mode=Time paced CTCV is preset.
4. Enter the Sampling interval.
5. Enter the Sampling volume per sample. (For version with vacuum pump, configure
under Menu/Setup/General settings/Sampling.)
6. Select the Bottle change mode after number of samples or time for average samples.
With the option "Bottle change after a time", you can enter the change time and bottle
synchronization (None, 1st bottle change time, 1st time of change + bottle number).
The description for this can be found in the "Bottle synchronization" section.
With the option "Bottle change after a time", you can choose the bottle
synchronization before the start condition (None, 1st bottle change time, 1st time of
change + bottle number). The description for this can be found in the "Bottle
synchronization" section.
1. For Multiple bottles enter the number of bottles the sample should be distributed
over.
2. Start condition: immediately or after date/time
3. Stop condition: after program end or continuous operation.
4. Pressing SAVE saves the program and ends data entry.