Liquistation CSF34 Operation
Endress+Hauser 71
Menu/Setup/(General settings or Inputs<Sensor channel>)/Extended setup/Diagnostics settings/Diag.
behavior
Function Options Info
Cleaning program Selection
• None
• Cleaning 1
• Cleaning 2
• Cleaning 3
• Cleaning 4
Factory setting
None
‣
Decide whether the diagnostic message
should trigger a cleaning program.
The cleaning programs can be defined under:
Menu/Setup/Additional functions/Cleaning.
Detail information Read only Additional information on the diagnostic
message and instructions on how to resolve the
problem.
HART bus address
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.
Menu/Setup/General settings/Extended setup/HART
Function Options Info
Bus address 0 to 63
Factory setting
0
You can change the device address to integrate
several HART devices in a single network
(Multidrop mode).
If you reset the device to the factory settings (Diagnostics/Reset/Factory default),
the bus address is not reset. Your setting is retained.
PROFIBUS DP
Menu/Setup/General settings/Extended setup/PROFIBUS
Function Options Info
Enable Selection
• Off
• On
Factory setting
On
You can switch off communication at this point.
The software can then only be accessed via local
operation.
Termination Read only If the device is the last in the bus, you can
terminate via the hardware.
Bus address 1 to 125 If you have addressed the bus via hardware (DIP
switches on the module, ), you can only read the
address here.
If an invalid address is set via the hardware, you
have to assign a valid address for your device
either here or via the bus.
Ident number Selection
• Automatic
• PA-Profile 3.02 (9760)
• Liquistation CSFxx
(155C)
• Manufacturer specific
Factory setting
Automatic