Operation Liquistation CSF34
86 Endress+Hauser
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Assignment bin. output Selection
• No binary output
configured
• Binary output Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Settings with the Basic program type with several bottles
Sampling mode = Time paced CTCV
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling interval
(for version with vacuum pump or
peristaltic pump)
00:01:00 to 99:59:00
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
00:10:00 HH:MM:SS
HH:MM:SS
Factory setting
00:10:00 HH:MM:SS
Set the sampling interval.
Dosing volume (for version with
vacuum pump or sampling
assembly)
Sampling volume (for version with
peristaltic pump)
Sampling volume
Factory setting
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling
volume.
Set the sampling volume.
In the version with a vacuum pump or
sampling assembly, the volume is taken
from the setup and can only be modified
there.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.
Multiplier
(for version with vacuum pump or
sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using
the multiplier. For example, if the dosing
volume is set to 200 ml, set the multiplier to
2 for a sampling volume of 400 ml. 2
samples are taken in succession when
sampling.
Bottle change mode Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
If you select Bottle change mode Number of samples