Liquistation CSF34 Operation
Endress+Hauser 87
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples.
If the bottle is full beforehand based on the
calculated level, the system prevents more
samples being added to the bottle. Such
samples are logged as failed samples in the
program logbook. At the same time, the
diagnostics message "Overfill check" (F353)
is also triggered. Set the number of samples.
If you select Bottle change mode Time
External event Selection
No bottle change input
configured
Binary input Sx
Factory setting
No bottle change input
configured
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Time interval 00-00:02 to 31-00:00
DD-HH:MM
Factory setting
0001:00 DDHH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Multiple bottles 0 to 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.
Bottle synchronization Selection
None
1. bottle change time
1. Time of change +
bottle number
Factory setting
None
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. Set the
synchronization time.
1. Time of change + bottle number
Each bottle is assigned to a specific fill time.
Set the synchronization time and the
weekday.
If you select Bottle change mode External signal
Bottle chg. sig. input Selection
No bottle change input
configured
Binary input Sx
Factory setting
No bottle change input
configured
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Multiple bottles 0 to 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.
Start condition Selection
Immediate
Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.
If you select Start condition Immediate