EasyManua.ls Logo

Endress+Hauser Liquistation CSF34 - Page 88

Endress+Hauser Liquistation CSF34
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Liquistation CSF34
88 Endress+Hauser
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sample at start Selection
Yes
No
Factory setting
Yes
Yes
The first sample is taken when the program
is started.
No
The system waits for the interval to elapse
before the first sample is taken.
Start date 01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Stop condition Selection
Program end
Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output Selection
No binary output
configured
Binary output Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Settings with a flow-paced Basic program
Settings with the Basic program type with 1 bottle
Sampling mode = Flow paced VTCV
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Flowmeter input Selection
No flow input
configured
Binary input S:x
Current input S:x
Factory setting
No flow input configured
Select the flow input. The binary input or the
current input must be configured for this
function.
Only the inputs configured as a flow input
are displayed.
Sampling interval (for version
with vacuum pump or peristaltic
pump)
Sampling interval
1000 to 9,999,000 m
3
3
Factory setting
10,000 m
3
3
Set the sampling interval. The unit and the
number of decimal places are displayed as
configured under Setup/ Inputs.

Table of Contents

Other manuals for Endress+Hauser Liquistation CSF34

Related product manuals