Operation Liquistation CSF34
90 Endress+Hauser
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Start date 01.01.2000 ... 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program.
The format depends on the option
configured under general settings.
Start time 00:00:00 ... 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is
started. The format depends on the option
configured under general settings.
Stop condition Selection
• Program end
• Continuous
Factory setting
Program end
Program end
The device stops sampling automatically
once it has run through the set program.
Continuous
The device runs through the set program
continuously in an infinite loop. Do not
forget to empty the bottles.
Assignment bin. output Selection
• No binary output
configured
• Binary output Sx
Factory setting
No binary output
configured
Assignment of the binary output to the
program cycle.
Inputs Settings for the inputs can be made as
described in the "Inputs" section.
Settings with the Basic program type with several bottles
Sampling mode = Flow paced VTCV
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Flowmeter input Selection
• No flow input
configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the flow input. The binary input or the
current input must be configured for this
function. Only the inputs configured as a
flow input are displayed.
Sampling interval (for version
with vacuum pump or peristaltic
pump)
Sampling interval
1000 to 9,999,000 m
3
3
Factory setting
10,000 m
3
3
Set the sampling interval. The unit and the
number of decimal places are displayed as
configured under Setup/ Inputs.
Dosing volume (for version with
vacuum pump or sampling
assembly)
Sampling volume (for version with
peristaltic pump)
Sampling volume
Factory setting
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling
volume.
Set the sampling volume.
In the version with a vacuum pump or
sampling assembly, the volume is taken
from the setup and can only be modified
there.
The dosing accuracy and the repeatability of
a sample volume < 20 ml can vary,
depending on the specific application.