Liquistation CSF34 Operation
Endress+Hauser 91
Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Multiplier
(for version with vacuum pump or
sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using
the multiplier. For example, if the dosing
volume is set to 200 ml, set the multiplier to
2 for a sampling volume of 400 ml. 2
samples are taken in succession when
sampling.
Bottle change mode Selection
• Number of samples
• Time
• External signal
Factory setting
External signal
The bottle can be changed after a specific
number of samples, after a time or by an
external signal.
If is selected: Bottle change mode Number of samples
Samples per bottle 1 ... 9999
Factory setting
1
Set the number of samples.
Having selected Bottle change mode Time
Time interval 00-00:02 ... 31-00:00
DD-HH:MM
Factory setting
00‐01:00 DD‐HH:MM
Set the time (days, hours and minutes) after
which the system should change to the next
bottle.
Multiple bottles 0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.
Bottle synchronization Selection
• None
• 1. bottle change time
• 1. Time of change +
bottle number
Factory setting
None
None
The sampling and bottle change times are
not synchronized.
1. bottle change time
Sampling starts with the first bottle. Set the
synchronization time.
1. Time of change + bottle number
Each bottle is assigned to a specific fill time.
Set the synchronization time and the
weekday.
Having selected Bottle change mode External signal
Bottle chg. sig. input Selection
• No bottle change input
configured
• Binary input Sx
Factory setting
No bottle change input
configured
The bottle change input can be configured
under Inputs .
A binary input must be configured for this
function. The sampling input can be
configured under "Inputs".
Multiple bottles 0 ... 23
The configuration options
depend on the current
number of bottles
Factory setting
0
Multiple bottles
"Simultaneous" transfer of two samples to
separate bottles.
Start condition Selection
• Immediate
• Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable
time.