The "Expert" menu Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus
120 Endress+Hauser
Options • Actual status
The switch output remains in the state it was in when the error occurred
• Open
• Closed
Factory setting Open
Switch status (0461)
Navigation
Expert → Output → Switch output → Switch status
Diagnostics → Measured value → Switch status
Description Indicates the current state of the switch output.
Display options • Open
• Closed
Invert output signal (0470)
Navigation Expert → Output → Switch output → Invert output signal
Description Allows to invert the behavior of the switch output.
Options • No
The behavior of the switch output is as described above.
• Yes
The states Opend and Closed are inverted as compared to the description above.