The "Expert" menu Micropilot FMR50/FMR51/FMR52/FMR53/FMR54/FMR56/FMR57 FOUNDATION Fieldbus
52 Endress+Hauser
Factory setting 0%
Additional information The value specified in this parameter is added to the measured level (before linearization).
Output mode (2317)
Navigation Expert → Sensor → Level → Output mode
Write access Maintenance
Description Defines the output mode
Options • Ullage
• Level linearized
Factory setting Level linearized
Additional information • Ullage
The unfilled space of the tank or silo is indicated.
• Level linearized
The level is indicated (more precisely: the linearized value if a linearization has been
activated).
A0016086
4 Definition of the "Output mode" parameter
A Level linearized
B Ullage
Level (2319)
Navigation Expert → Sensor → Level → Level
Write access Read-only parameter