Proline Promag 10 Installation
Endress+Hauser 11
Inlet and outlet runs
If possible, install the sensor upstream from fittings such as valves, T-pieces, elbows, etc.
2.2.4 Vibrations
Secure and fix both the piping and the sensor if vibrations are severe.
A0008160
The following inlet and outlet runs must be observed
in order to meet accuracy specifications:
Inlet run: 5 × DN
•Outlet run: 2 × DN
A0008161
Measures to prevent device vibration,
L > 10 m (33 ft)
"
Caution!
It is advisable to install the sensor and transmitter
separately if vibration is excessively severe. For
information on the permitted shock and vibration
resistance , see the Operating Instructions on the
CD-ROM.
5xDN³
³ 2xDN