Design Proline Promag 300 HART
12 Endress+Hauser
Feature Designation Option selected
480 Device model All
500 Display operating language All
520 Sensor option All
530 Customer-specific configuration All
540 Application package EA "Extended HistoROM"
EB "Heartbeat Verification + Monitoring"
EL "OPC-UA server"
550
2)
Suitability for custody transfer measurement All
570 Service All
580 Test, certificate All
590 Additional approval LA (= SIL)
3)
610 Accessory mounted All
620 Accessory enclosed All
850 Firmware version Firmware with SIL capability, e.g. 01.01.zz (HART)
895 Marking All
1) In devices with several outputs, only current output 1 (terminals 26 and 27) is suitable for safety functions. The
other outputs can, if necessary, be connected for non-safety-oriented purposes.
2) Only for devices with approval for custody transfer
3) Additional selection of further approvals is possible.
4.1.2 Suitability of the measuring device
1. Carefully select the nominal diameter of the measuring device in accordance with the
application's expected flow rates.
The maximum flow rate during operation must not exceed the specified maximum
value for the sensor.
For further information on selecting the nominal diameter, see the Technical
Information
2. In SIL mode: ensure that the measuring tube is filled.
Empty pipe detection is switched to the Off option in the SIL mode.