Performance characteristics
Reference operating
conditions
• Error limits following DIN EN 29104, future ISO 20456
• Water, typically +15 to +45°C (+59 to +113 °F); 0,5 to 7 bar (73 to 101 psi)
• Specification as per calibration protocol
• Data on the measured error based on accredited calibration rigs traced back to ISO 17025
Maximum measured error Promag 50
• Pulse output: ±0.5% o.r. ± 1 mm/s
optional: ±0.2% o.r. ± 2 mm/s
(o.r. = of reading)
• Current output: also typically ± 5 A
Promag 53
• Pulse output: ±0.2% o.r. ± 2 mm/s (o.r. = of reading)
• Current output: also typically ± 5 A
Fluctuations in the supply voltage do not have any effect within the specified range.
A0005531
Max. measured error in % of reading
Repeatability Max. ±0.1% o.r. ± 0.5 mm/s (o.r. = of reading)