14 Endress+Hauser
Performance characteristics
Reference operating
conditions
Error limits following DIN EN 29104, future ISO 20456
Water, typically +15 to +45°C (+59 to +113 °F); 0,5 to 7 bar (73 to 101 psi)
Specification as per calibration protocol
Data on the measured error based on accredited calibration rigs traced back to ISO 17025
Maximum measured error Promag 50
Pulse output: ±0.5% o.r. ± 1 mm/s
optional: ±0.2% o.r. ± 2 mm/s
(o.r. = of reading)
Current output: also typically ± 5 A
Promag 53
Pulse output: ±0.2% o.r. ± 2 mm/s (o.r. = of reading)
Current output: also typically ± 5 A
Fluctuations in the supply voltage do not have any effect within the specified range.
A0005531
Max. measured error in % of reading
Repeatability Max. ±0.1% o.r. ± 0.5 mm/s (o.r. = of reading)
2.5
[%]
2.0
1.5
1.0
0.5
0
0.2 %
0.5 %
0
1
2 4 6 8 10 [m/s]
v
5 10 15 20 25 30 32 [ft/s]0