Proline Promag 50H, 53H
Endress+Hauser 19
A0016359
Pressure loss due to adapters
Length of connecting cable When mounting the remote version, please note the following to achieve correct measuring results:
Fix the cable run or route it in an armored conduit. Cable movements can falsify the measuring signal
especially in the case of low fluid conductivities.
Route the cable well clear of electrical machines and switching elements.
If necessary, ensure potential equalization between sensor and transmitter.
The permissible cable length L
max
depends on the fluid conductivity. A minimum conductivity of
20 S/cm is required for measuring demineralized water.
When the empty pipe detection function is switched on (EPD), the maximum connecting cable
length is 10 m (33 ft).
A0010734
Permitted length of connecting cable for remote version
Area marked in gray = permitted range; L
max
= length of connecting cable in [m] ([ft]); fluid conductivity in [μS/cm]
100
10
0.5
d / D
[mbar]
0.6 0.7 0.8 0.9
1 m/s
2 m/s
3 m/s
4 m/s
5 m/s
6 m/s
7 m/s
8 m/s
1
D
d
max. 8°
200
100
5
10 100 200
L
max
[m][m]
[µS/cm]
L
max
[ft]
200 6000
400