Installation
Mounting location Entrained air or gas bubble formation in the measuring tube can result in an increase in measuring
errors.
Avoid the following installation locations in the pipe:
• Highest point of a pipeline. Risk of air entrainment.
• Directly upstream from a free pipe outlet in a vertical pipeline.
A0011899
Mounting location
Installation of pumps
Sensors may not be installed on the pump suction side. This precaution is to avoid low pressure and the
consequent risk of damage to the lining of the measuring tube. Information on the pressure tightness
of the measuring tube lining 28, "Pressure tightness" section.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or
peristaltic pumps. Information on the shock and vibration resistance of the measuring system 20,
"Shock and vibration resistance" section.
A0011900
Installation of pumps