Proline Promag 50H, 53H
Endress+Hauser 9
Supply voltage 85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
PROFIBUS PA and FOUNDATION Fieldbus
Non-Ex: 9 to 32 V DC
Ex i: 9 to 24 V DC
Ex d: 9 to 32 V DC
Power consumption AC: < 15 VA (incl. sensor)
DC: < 15 W (incl. sensor)
Switch-on current:
Max. 3.0 A (< 5 ms) for 260 V AC
Max. 8.5 A (< 5 ms) for 24 V DC
Power supply failure Lasting at least one half cycle frequency: EEPROM saves measuring system data
EEPROM or T-DAT (Promag 53 only) retain the measuring system data in the event of a power
supply failure
S-DAT: exchangeable data storage chip which stores the data of the sensor (nominal diameter, serial
number, calibration factor, zero point etc.)