Proline Promag 50H, 53H
6 Endress+Hauser
Promag 53
Current output
active/passive selectable, galvanically isolated, time constant selectable (0.01 to 100 s), full scale value
selectable, temperature coefficient: typ. 0.005% o.r./°C (o.r. = of reading), resolution: 0.5 mA
active: 0/4 to 20 mA, R
L
< 700 (for HART: R
L
250 )
passive: 4 to 20 mA; supply voltage V
S
: 18 to 30 V DC; R
i
150
Pulse/frequency output
active/passive selectable, galvanically isolated (Ex i version: only passive)
active: 24 V DC, 25 mA (max. 250 mA during 20 ms), R
L
100
passive: open collector, 30 V DC, 250 mA
Frequency output: full scale frequency 2 to 10000 Hz (f
max
= 12500 Hz), for EEx-ia 2 to 5000 Hz;
on/off ratio 1:1, pulse width max. 10 s
Pulse output: pulse value and pulse polarity selectable, max. pulse width configurable (0.05 to 2000
ms)
PROFIBUS DP interface
Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A: 1998,
galvanically isolated
Profile version 3.0
Data transmission rate: 9.6 kBaud to 12 MBaud
Automatic data transmission rate recognition
Function blocks: 2 × analog input, 3 × totalizer
Output data: volume flow, calculated mass flow, totalizer 1 to 3
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device
Available output combination 8
PROFIBUS PA interface
Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated
Profile version 3.0
Current consumption: 11 mA
Permissible supply voltage: 9 to 32 V
Bus connection with integrated reverse polarity protection
Error current FDE (Fault Disconnection Electronic): 0 mA
Function blocks: 2 × analog input, 3 × totalizer
Output data: volume flow, calculated mass flow, totalizer 1 to 3
Input data: positive zero return (ON/OFF), totalizer control, value for local display
Cyclic data transmission compatible with previous model Promag 33
Bus address adjustable via miniature switches or local display (optional) at the measuring device
Modbus RS485 interface
Transmission technology (Physical Layer): RS485 in accordance with ASME/TIA/EIA-485-A: 1998,
galvanically isolated
Modbus device type: slave
Address range: 1 to 247
Bus address adjustable via miniature switches or local display (optional) at the measuring device
Supported Modbus function codes: 03, 04, 06, 08, 16, 23
Broadcast: supported with the function codes 06, 16, 23
Transmission mode: RTU or ASCII
Supported baudrate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
•Response time:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
Available output combination 8