Proline Promag W 400
Endress+Hauser 15
Signal transmission 0-20 mA/4-20 mA HART with additional outputs and inputs
Order code for
"Output" and
"Input"
Terminal numbers
Output 1 Output 2 Output 3 Input
26 (+) 27 (-) 24 (+) 25 (-) 22 (+) 23 (-) 20 (+) 21 (-)
Option H • 4-20 mA HART
(active)
• 0-20 mA
(active)
Pulse/frequency
output
(passive)
Switch output
(passive)
-
Option I • 4-20 mA HART
(active)
• 0-20 mA
(active)
Pulse/frequency/
switch output
(passive)
Pulse/frequency/
switch output
(passive)
Status input
Option J • 4-20 mA HART
(active)
• 0-20 mA
(active)
Permanently
assigned
Pulse output
adjusted
(passive)
Pulse/frequency/
switch output
(passive)
Status input
EtherNet/IP connection version
The sensor can be ordered with terminals or a device plug.
Connection methods available
Possible options for order code
"Electrical connection"
Outputs
Power
supply
Terminals Terminals • Option A: coupling M20x1
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
Device plug
(→ 16)
Terminals • Option L: plug M12x1 + thread NPT ½"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
A0020428
1 Supply voltage (wide range power unit)
2 EtherNet/IP