Proline Promag W 400
32 Endress+Hauser
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
Complies with emission limits for industry as per EN 55011 (Class A)
Details are provided in the Declaration of Conformity.
Process
Medium temperature range
0 to +80 °C (+32 to +176 °F) for hard rubber, DN 50 to 2000 (2 to 78")
–20 to +50 °C (–4 to +122 °F) for polyurethane, DN 25 to 1200 (1 to 48")
Conductivity
≥ 5 μS/cm for liquids in general
Note that in the case of the remote version, the requisite minimum conductivity also depends
on the cable length (→  29).
Pressure-temperature
ratings
The following material load diagrams refer to the entire device and not just the process connection.
Flange connection according to EN 1092-1 (DIN 2501)
0
5
10
15
20
25
35
30
40
[bar][psi]
-40
-20
0 20
40 60 80 100120140160180 [°C]
360 [°F]
0-40 100 200 300
200
100
400
300
500
600
0
PN25
PN16
PN10
PN 6
PN40
A0021188-EN
 19 Fixed flange PN 6/10/16/25/40, materials C22, FE 410W B and S235JRG2, DN 25 to 2000 (1 to 78")