EasyManua.ls Logo

Endress+Hauser Proline Promag W 400 - Electrical Connection; Connection Conditions

Endress+Hauser Proline Promag W 400
92 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Promag W 400
36 Endress+Hauser
A0015594
Vibrations
In the event of very strong vibrations, the pipe and sensor must be supported and fixed.
It is also advisable to mount the sensor and transmitter separately.
For information on the permitted resistance to vibration and shock (→  31), (→  31)
L
A0016266
 26 Measures to prevent vibration of the device
L > 10 m (33 ft)
Custody transfer measurement
As an option, Promag W 400 is tested in accordance with OIML R49 and has an EC type-
examination certificate according to Measuring Instruments Directive 2004/22/EC (MID) for service
subject to legal metrological control ("custody transfer") for cold water (Annex MI-001).
The permitted fluid temperature in these applications is 0 to +50 °C (+32 to +122 °F).
The device is used with a legally controlled totalizer on the local display and optionally with a legally
controlled pulse output.
Measuring devices subject to legal metrological control totalize in both directions, i.e. all the outputs
consider flow components in the positive (forward) and negative (reverse) flow direction.
Generally a measuring device subject to legal metrological control is secured against tampering by
seals on the transmitter or sensor. These seals may normally only be opened by a representative of
the competent authority for legal metrology controls.
After putting the device into circulation or after sealing the device, operation is only possible
to a limited extent.
Detailed ordering information is available from your local Endress+Hauser sales center for
national approvals as cold water meters based on OIML R49.