Potential equalization Requirements
Please consider the following to ensure correct measurement:
Same electrical potential for the fluid and sensor
Company-internal grounding concepts
Pipe material and grounding
Connection examples for standard situations
Metal, grounded pipe
A0016315
 9 Potential equalization via measuring tube
Connection example in special situations
Unlined and ungrounded metal pipe
This connection method also applies in situations where:
The customary potential equalization is not used
Equalizing currents are present
DN 300£ DN 350³
A0016317
 10 Potential equalization via ground terminal and pipe flanges
Note the following when installing:
Connect both sensor flanges to the pipe flange via a ground cable and ground them.
Connect the connection housing of the transmitter or sensor to ground potential by means of the
ground terminal provided for the purpose.To mount the ground cable:
If DN ≤ 300 (12"): Mount the ground cable directly on the conductive flange coating of the
sensor with the flange screws.
If DN ≥ 350 (14"): Mount the ground cable directly on the metal transport bracket.
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
For remote device versions, the ground terminal in the example always refers to the sensor and
not to the transmitter.