Proline Promag W 400
Endress+Hauser 35
Flange connection according to JIS B2220
10K
20K
-40
-20
0 20
40 60 80 100120140 160180 [°C]
0
10
20
30
[bar]
360 [°F]
0-40 100 200 300
[psi]
200
100
400
300
0
A0021183-EN
25 Fixed flange 10K/20K, materials HII, S235JRG2 and 1.0425 (316L), DN 25 to 300 (1 to 12")
Pressure tightness
Liner: hard rubber
Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
[mm] [in] +25 °C (+77 °F) +50 °C (+122 °F) +80 °C (+176 °F)
50…2000 2…78 0 (0) 0 (0) 0 (0)
Liner: polyurethane
Nominal diameter Limit values for absolute pressure in [mbar] ([psi]) for fluid temperatures:
[mm] [in] +25 °C (+77 °F) +50 °C (+122 °F)
25…1200 1…48 0 (0) 0 (0)
Flow limit
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum flow velocity is between 2 to 3 m/s (6.56 to 9.84 ft/s). Also match the velocity of flow (v)
to the physical properties of the fluid:
• v < 2 m/s (6.56 ft/s): for abrasive fluids (e.g. potter's clay, lime milk, ore slurry)
• v > 2 m/s (6.56 ft/s): for fluids producing buildup (e.g. wastewater sludges)
For an overview of the measuring range full scale values, see the "Measuring range" section
(→ 6)
For custody transfer, the applicable approval determines the permitted measuring range.
Pressure loss
• No pressure loss occurs if the sensor is installed in a pipe with the same nominal diameter.
• Pressure losses for configurations incorporating adapters according to DIN EN 545 (→ 28)
System pressure
• Never install the sensor on the pump suction side in order to avoid the risk of low pressure, and
thus damage to the liner.
• Furthermore, install pulse dampers if reciprocating, diaphragm or peristaltic pumps are used.
• For information on the liner's resistance to partial vacuum (→ 35)
• For information on the measuring system's resistance to vibration and shock (→ 31),
(→ 31)