15 Group SYSTEM PARAMETER Device Functions Proline Promass 80
56 Endress+Hauser
DENSITY DAMPING The density filter allows the sensitivity of the density measuring signal to be lowered with
respect to variations in the density of the fluid, e.g. with inhomogeneous liquids.
User input:
max. 5-digit number, including unit: 0.00...100.00 s
Factory setting:
0.00 s
! Note!
The damping acts on all functions and outputs of the measuring device.
FLOW DAMPING Use this function to set the filter depth of the digital filter. Using the interference blanking
(= time constant for exponential filter) the sensitivity of the flow measurement signal can
be reduced with respect to transient flows and interference peaks; e.g. with fluid
containing solids or gas bubbles, etc. The reaction time of the measuring system increases
with an increase in the filter setting.
User input:
0...100 s
Factory setting:
0 s
! Note!
The damping acts on all functions and outputs of the measuring device.