11 Group PULSE/FREQUENCY OUTPUT Device Functions Proline t-mass 65
42 Endress+Hauser
SIMULATION
SWITCH POINT
Use this function to activate simulation of the status output.
! Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function.
Options:
OFF
ON
Factory setting:
OFF
! Note!
• The “SIMULATION STATUS OUTPUT” message indicates that simulation is active.
• The flowmeter continues to measure while simulation is in progress, i.e. the current
measured values are output correctly via the other outputs.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
SWITCH POINT
Use this function to define the switching response of the status output during the simula-
tion. This value is used to test downstream devices and the flowmeter itself.
! Note!
This function is not available unless STATUS setting was selected in the OPERATION
MODE function and the ON setting was selected in the SIMULATION SWITCH POINT
function.
User input:
NOT CONDUCTIVE
CONDUCTIVE
Factory setting:
NOT CONDUCTIVE
! Note!
If exiting this function and the SIMULATION SWITCH POINT function is active (= ON),
the message END SIMULATION is displayed. Select YES to end simulation or NO to con-
tinue simulation.
" Caution!
The setting is not saved if the power supply fails.
Function description PULSE/FREQUENCY OUTPUT