11 Group PULSE/FREQUENCY OUTPUT Device Functions Proline t-mass 65
42 Endress+Hauser
SIMULATION
SWITCH POINT
Use this function to activate simulation of the status output.
! Note!
This function is not available unless the STATUS setting was selected in the OPERATION
MODE function.
Options:
OFF
ON
Factory setting:
OFF
! Note!
The “SIMULATION STATUS OUTPUT” message indicates that simulation is active.
The flowmeter continues to measure while simulation is in progress, i.e. the current
measured values are output correctly via the other outputs.
" Caution!
The setting is not saved if the power supply fails.
VALUE SIMULATION
SWITCH POINT
Use this function to define the switching response of the status output during the simula-
tion. This value is used to test downstream devices and the flowmeter itself.
! Note!
This function is not available unless STATUS setting was selected in the OPERATION
MODE function and the ON setting was selected in the SIMULATION SWITCH POINT
function.
User input:
NOT CONDUCTIVE
CONDUCTIVE
Factory setting:
NOT CONDUCTIVE
! Note!
If exiting this function and the SIMULATION SWITCH POINT function is active (= ON),
the message END SIMULATION is displayed. Select YES to end simulation or NO to con-
tinue simulation.
" Caution!
The setting is not saved if the power supply fails.
Function description PULSE/FREQUENCY OUTPUT