WARNING
Some circuit and
accessories configurations (mainly in leak
pediatric configuration) with
high resistive pressure in the circuit could lead to ineffective "Disconnection alarm".
For ventilator dependent patient, "Disconnection alarm" must be tested after any
calibration, setting changes or circuit configuration change.
In case the disconnection alarm detection is not efficient, it is mandatory to set a VTI Min
alarm (leak configurations) or a VTI Max alarm (valve configurations) as a backup for
disconnection events covering.