Pres. Control.
(Pressure Controlled -
mb)
Sets the pressure control above PEEP delivered during inspiration for
pressure-controlled breaths (with constant I Time).
P. Contr. Max
(Pressure Controlled
max - mb)
Sets the maximum pressure control above
PEEP delivered during
inspiration for pressure-controlled breaths (with constant I Time).
PSV
Pressure support Ventilation (with expiration valve)
PSV VT
Pressure support ventilation Volume regulated (with expiratory valve)
P
-
SIMV
Pressure Synchronous intermittent Mandatory Ventilation (with
expiratory valve)
Rate (C/min)
Sets the breaths per minute (bpm) to be delivered by the ventilator to
the patient. The measured respiratory rate may be greater due to
manual breaths.
S
T
Synchronized Timed (with leak)
I Time (s)
Sets the duration of the inspiratory phase of a breath.
I Time Max
(Maximum inspiratory
time - s)
Sets the maximum duration of the inspiratory phase of a breath. AUTO
/ 0.3-2.5
I Time Min
(Minimum
Inspiratory Time - s)
Sets the minimum duration of the inspiratory phase of a breath.
Inspiration Trigger (I
Trigger)
Sets the trigger threshold to be detected to deliver a new breath.
The trigger is blocked for the first 300 ms (Pediatric) or 500 ms (Adult)
following the beginning of expiration phase.
V
-
SIMV
Volume Synchronous intermittent Mandatory Ventilation (with
expiratory Valve)
VT
(Tidal Volume - ml)
Sets the volume of gas, to be delivered to the patient in during the
inspiratory phase of controlled volume breath in (A)VCV, MPV, PSV VT
or VTS modes.
VTS
Volume Target Synchronised (with leak)
Measured and calculated parameter definitions
These measured and calculated parameters are displayed during configuration or during ventilation.
Each ventilation mode sets out the parameters displayed.
FiO2 %
Fraction of inspired Oxygen delivered to Patient circuit.
I / E
The time ratio of the inspiratory period to the expiratory period. The
measured I / E ratio is displayed as a monitored parameter during
ventilation. Relative I / E ratio value is calculated and displayed during
settings of time variables like Respiratory rate or Ti.
Leak Level (l/min)
Level of intentional leak set in the patient configuration menu. Leak
monitoring displays leak above this set level.
MV
(Delivered Minute
Volume)
The measured volume averaged over the last breath. The measured MV is
displayed as a monitored parameter during ventilation