EasyManua.ls Logo

eove EO-150 - Connecting Two Power Sources Using the y Cable (EO-CPLPACK or EO-CPLPACKBOX)

eove EO-150
85 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50
Connecting two power sources using the Y cable (EO-CPLPACK or EO-CPLPACKBOX):
A Y cable is available to secure ventilator dependent patients in mobility or when the AC power
connection is unsafe. The solution is based on Battery Pack (EO-BAT9) accessory.
Refer to the Battery Pack (EO-BAT9) user manual for more details on the Battery Pack usage.
Connecting a Y cable with Battery Pack (EO-BAT9) and AC Power/Charger (EO-YCBLPWR):
WARNING
When using the Y cable solution with (EO
-
BAT9) and AC Power/Charger (EO
-
YCBLPWR) or when using EO-CPLPACKBOX cable, The Battery Pack (EO-BAT9) needs
to be charged at least every two weeks with its own charger. Check regularly the
Battery Pack autonomy level indicated on the Battery Pack keypad.
CAUTION
The shorter part of the Y cable needs to be connected to the primary power source (AC
Power/Charger)
NOTE:
In case of AC power loss combined with a battery failure while using the Y cable
solution with (EO-BAT9) and AC Power/Charger (EO-YCBLPWR), the ventilator can reset
and restart ventilation normally within 5 seconds.

Table of Contents