63
Servicing
Maintenance of the ventilator should be carried out by a trained technician. Attempting to
repair the machine yourself could result in patient injury or damage to the machine.
It is forbidden to modify the EO 150
without manufacturer authorization.
Retain the original packaging to use when shipping to/from service agent.
Maintenance Timetable
The EO-150 should be regularly serviced by an authorized EOVE technician according to the
following schedule. The ventilator will provide safe and reliable ventilation for 10 years provided
that it is operated and maintained in accordance with the instructions given in this manual. As with
all electrical devices, if any problem arises with your EO 150 device, you should exercise caution and
have it inspected by an authorized EOVE technician.
Servicing schedule from the date of first use:
Person trained in the use of
EO-150
Check the air filter and replace
if necessary (replace earlier if
dirty or dusty).
Check Double limb circuit
adapter membrane, if used.
Replace if necessary.
Qualified EOVE technician
service notification is
displayed.
Qualified EOVE technician
Replace turbine depending on
the settings chosen and the
patient profile (adult or
pediatric)