41
To attach an antibacterial filter to the EO-150:
1. Attach the antibacterial filter to the inspiratory port of the device.
2. Connect the breathing tube to the other side of the filter.
3. Perform a calibration.
4. Connect the patient interface to the other end of the breathing tube.
Attaching a humidifier
Humidification of the inspired gas is required for invasive ventilation in order to prevent
any lung injury.
humidifier on a level surface lower than level of the ventilator and the
patient in order to prevent the mask and tubing filling with water.
Ensure that the humidifier is set up according to the manufacturer's instructions.
Use appropriate precautions
to prevent water in the circuit transferring to the patient (e.g.,
a water trap).
Only use HMEs that comply with the relevant safety standards, including ISO 9360
ISO 9360-2.
Make sure that the water tub is empty and
thoroughly dried before transporting the
humidifier.
To attach a humidifier to a circuit configuration:
1. Connect the air tubing to the inspiratory port on the device.
2. Connect the other end of the air tubing to the inlet port on the humidifier.
3. Connect the patient circuit to the outlet port on the humidifier.