84
Pressure
The measured airway pressure at the patient inlet po
rt. The measured
Pressure is displayed in the monitoring menu during ventilation
PEEP (Positive End
Expiratory pressure
-mb)
The airway pressure measured 100 ms at the end of the last expiration. The
measured PEEP is displayed in the monitoring menu during ventilation.
Pmean
Mean airway pressure in the Patient circuit computed over the last breath.
PIP (Peak
Inspiratory
Pressure -mb)
The maximum airway pressure reached over the last inspiration. The
measured PIP is displayed in the monitoring menu during ventilation
Heart rate
When a pulse oximeter is connected on the ventilator Pulse rate is
displayed in the monitoring menu during ventilation.
Total Breath Rate
(RR) (Res. Rate -
c/min)
Breath rate is computed over the last
breath according to (60/inspiratory
time + Expiratory time). The measured Rtot is displayed in the monitoring
menu during ventilation.
SpO2
(
Measured
Oxygen Saturation
%)
When a pulse oximeter is connected on the ventilator, SpO2 is displayed in
the monitoring menu during ventilation.
Te (Expiratory time
-s)
The duration in
seconds of the last expiratory phase. The measured Te is
displayed in the monitoring menu during ventilation.
Ti (Inspiratory time
-s)
The duration in seconds of the last inspiratory phase. The measured Ti is
displayed in the monitoring menu during ventilation.
Exhalation Tidal
Volume (VTE -ml)
The expired volume measured over the last breath. Vte is displayed in the
monitoring menu during ventilation.
Inspiratory Tidal
Volume (VTI- ml)
The inspired volume measured over the last breath. Vti is
displayed in the
monitoring menu during ventilation.
Other Definitions
IP22
Ingress protection level
IP2X
protects against the ingress of solid objects with >12.5 mm diameter.(e.g. fingers)
IPX2
protects against the ingress of water droplets (at 15° angle tilt)