EasyManua.ls Logo

eove EO-150 - Power Specifications; Environmental Specifications

eove EO-150
85 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
74
Power specifications
WARNING
This device is intended to function with external power supply 2440 from Mascot, never
use any other power supply unless recommended by Eove.
To disconnect the device from the mains, unplug power supply.
AC Inlet Voltage
100
-
240V
AC Inlet Power
1.6 A max
AC Inlet Power
50
-
60 Hz
DC inlet voltage
Max current
12 to 30 V
2.5 to 5.8 A
Power
105 W maximum (peaks)
70 W nominal
Module Embedded battery life
5 hrs
Internal battery (not replaceable by user) :
capacity
voltage
discharge current
Lithium
-
ion
2,8 Ah
21,6 V nominal
7 A max.
Interface/touchscreen start up time
< 50
s
Ventilator unit start up time
5 seconds
Environmental Specifications
Storage and transport conditions:
Ambient temperature
From
-
20°C à +60°C.
Relative humidity
From 10% à 95%, (non
-
condensing)
Operating conditions:
Ambient temperature
From +5°C to +40°C (after conditioning at 23° for 20
minutes)
Relative humidity
From 10% à 95%, (non
-
condensing).
Atmospheric pressure
From 660 hPa à 1100 hPa. (by default EO150
compensates for atmospheric pressure variations
e.g. related to altitude up to 3000 m).
O2 pressure source
O2 pressure source: up to
50
K
pa
with fl
ow up to
20 l/min and flexible tubing used. (from hospital O2
network, always use a flow or pressure limiter)
Bluetooth maximum distance
4 to 8 m

Table of Contents