47
Battery run time
When the internal battery is being used to power the device, the amount of charge remaining in the
battery is displayed as shown in the following table.
Touch
Screen
Keyboard
Display Description
When the internal battery is in use, the battery charge level is displayed
by percentage on the touch screen and by 4 LEDs on the keyboard.
When the internal battery is charging, the charge battery symbol is
displayed on the touch screen and by scrolling LEDs on the keyboard.
When the internal battery is charging but
may be
not enough charged to
supply the device, the yellow charge battery symbol is displayed on the
touch screen and by scrolling LEDs on the keyboard (also displayed in case
of some battery failure alarm).
When the internal battery power is low, the charge battery symbol is
displayed with an exclamation mark on the touch screen and the LED’s on
the keyboard are red.
Alarms will alert the user when the battery power is getting low.
Internal battery run time is determined by:
Environmental conditions (operating conditions See Technical Specifications)
The condition and age of the battery
The device settings
The current circuit in place and unintentional leak
The ventilation module internal battery will operate for approximately 5 hours (+/-10%) when the
device is used according to the following configuration for an adult patient (Resistance: 20 mb/l/s,
Compliance: 25 ml/mb): Inspiratory pressure: 20 cmH2O, Ventilation Rate: 15 bpm, I/E ratio: 1/2,
PEEP: OFF
The ventilation module internal battery will operate for approximately 4,5 hours (+/-10%) when the
device is used according to the following configuration for an adult patient (Resistance: 5 mb/l/s,
Compliance: 50 ml/mb): Inspiratory volume: 800 ml, Ventilation Rate: 20 bpm, I/E ratio: 1/2
The ventilation module internal battery will operate for approximately 3,5 hours when the
device is used according to the following configuration for an adult patient (Resistance: 20
mb/l/s, Compliance: 25 ml/mb): Inspiratory pressure: 30 cmH2O, Ventilation Rate: 15 bpm, I/E
ratio: ½, PEEP: 10 mb
WARNING
Those reference
internal
battery
run
time are given for a ventilation module
without
docking station. When used with docking station, the resulting operation time will
depend on luminosity setting and stand-by mode option and can be reduced by 20% to
30%. Always look at the device indication to know the real conditions expected
autonomy.