Monitored Parameter Specifications
(Rounded values for readings)
Parameter
Range
Display method
/ Filtering
Peak Inspiratory Pressure (PIP)
0 to 99 mbar
Computed
for
each
inspiration
Positive End Expiratory Pressure (PEEP)
0 to 60 mbar
Computed for each exhalation
Inspiratory Tidal Volume (VTI)
0 to 4000 ml
Computed for each inspiration
Exhalation Tidal Volume (VTE)
0 to 4000 ml
Computed for each exhalation
Total Breath Rate (Rtot)
1
to 99
.9
bpm
Computed for each breath
I E Ratio (I E)
33.3
:1
to 1:
9
9
.9
Computed for each breath
Inspiratory Time (I Time)
0.
1
to 9.9 s
Computed for each inspiration
Exhalation Time (E Time)
0
.3
to 59.9 s
Computed for each exhalation
I Peak Flow
0 to 100 l/min
Computed for each inspiration
E Peak Flow
0 to
100 l/min
Computed for each exhalation
Inspiratory Minute Volume (Min VI)
0 to
99.9
l
Computed for each breath
Leaks
0 to 200 l/min
Computed for each breath
FiO2
21
%
to 100%
Computed for each breath
SpO2
80%
to 100
%
According to NONIN or
Sentec
Heart Rate
30
-
250 (Sentec)
18-300 (NONIN)
According to NONIN or Sentec
PCO2
0
-
200
According to Sentec
A monitored value displayed as “--“ means that the measurement is not available or invalid.
Accuracy of monitoring data
Inspired volume in valve modes : ± (5 ml + 10%)
Expired volume in double limb circuit : ± (5 ml + 20%)
Leak volume : ± (10 ml + 10%)
Pressure : ± (2 mb + 4%)
Time : ± 0.1 s
I/E : Calculation based on Inspiratory Time and Exhalation Time
Rate : ± 1 breath/min
Flow : ± (1 l/min + 10%)
Peak flow : ± (2 l/min + 15%)
FIO2 : ± 3%
SPO2 : According to NONIN or Sentec specifications
Heart Rate : According to NONIN or Sentec specifications
PCO2 : According to Sentec specifications