Alarm Parameter Specifications
Alarm sound level: 55 - 75 dB ± 10%
The ventilator has the following alarm settings in specific ventilation modes:
Settings
Adult
Pediatric
Modes
Pres. min (mb) *
2
-
55
5
-
55
(A)VCV, MPV, VSIMV
, C
-
FLOW
Pres. max (mb) *
10
-
80
10
-
80
(A)VCV, MPV, VSIMV
VTI Min (ml)
50
-
2490
20
-
600
All except
(A)VCV, MPV, MPP
, C
-
FLOW
VTI Max (ml)
60
-
3000
40
-
800
(A)PCV, PSV, PSV VT, VSIMV, PSIMV
VTE Min (ml)
50
-
2490
20
-
600
All valve modes
VTE Max (ml)
60
-
3000
40
-
800
All valve modes
MV Min
1
-
25
0.5
-
6
All except
(A)VCV
, C
-
FLOW
Rate Min
6
-
65
6
-
85
PSV, PSV VT,
ST, VTS, CPAP
Rate Max
10
-
70
20
-
90
All except MPV, MPP
, C
PAP, C
-
FLOW
FIO2 Min
18
-
80
18
-
80
All
FIO2 Max
30
-
100
30
-
100
All
SPO2 Min
80
-
95
80
-
95
All
Disc. Time
AUTO / 5
-
120
AUTO / 5
-
60
All except MPV, MPP, C
-
FLOW
Disc. Time
OFF / 5
-
120
OFF / 5
-
60
C
-
FLOW
Disc. Time
OFF / 5
-
900
OFF / 5
-
900
MPV, MPP
*PWmax and PWmin according to ISO 10651
-
2 / ISO 80601
-
2
-
72
WARNING
Delaying or deactivating the disconnection alarm must be done only with patients that
have the corresponding capacity of breathing spontaneously. Failure to apply this
recommendation can lead to a life-threatening risk for the patient.