Accessories Compatible with EO-150
The EO-150 ventilator is compatible with a range of accessories.
AC Power/Charger Mascot 2440 (EO-PWRCHRG)
FIO2 Cable (O2CELCBL)
Car Adaptor DC Cable (EO-CARCBL)
SPO2 Cable (EO-SPO2CBL)
Transport Bag (EO-CARBAG1X0)
Nomad Bag (EO-NOMADBAG)
Travel Bag (EO-TRVELBAG1X0)
Remote Alarm Cable (EO-ALARMCBL)
Proximal Flow sensor (EO-PFLOWS)
Battery Pack (EO-BAT9)
Y cable (EO-CPLPACK)
AC Power/Charger for Y cable (EO-YCBLPWR)
Trolley (KC072283*)
Adjustable arm for trolley (KB029900*)
* From Air Liquide Medical Systems S.A. Parc de Haute Technologie 6 rue Georges Besse
92182 ANTONY CEDEX – France)
WARNING
Before using any accessory, always carefully read the accompanying Quick User Guide and
User Manual.
The EO
-
150 Ventilator should only be used with accessories recommended by EOVE.
Connection of other accessories could result in patient injury or damage to the device.
Attaching patient circuit accessories
WARNING
Adding or removing circuit components can adversely affect ventilation performance.
A circuit calibration is recommended every time an accessory or component is added to
or removed from the patient circuit.
Do not use electrically conductive or anti
-
static air tubing.
Attaching an antibacterial filter
WARNING
To prevent the risk of cross
-
contamination, an antibacterial filter is mandatory if the device
is to be used on multiple patients.
Regularly check the antibacterial filter and expiratory valve for signs of moisture or other
contaminants, particularly during nebulisation or humidification. Failure to do so could
result in increased breathing system resistance and/or inaccuracies in expired gas
measurement.
Only use antibacterial filters that comply with the relevant safety standards, including ISO
23328-1 and ISO 23328-2.
CAUTION
The antibacterial filter must be used and replaced
according to the manufacturer's
specifications.