EX-2001 NC RACER ZSME300000091
16
EXCELL PRECISION CO., LTD.
Before linearity calibration, make sure “Zero calibration” and “Weight Calibration” are properly
performed. Set the calibration switch to ON position.
Press key
Press key
Choose the calibration value Press key to previous step
Press key
Press key Press key to previous step
Enter the correct weight value
Press key
The calibrated weight value
is displayed when the
signal is stable
Press to next step
After finishing the first linearity calibration,
choose to continue or exit the linearity mode
by pressing .
Refer to 5-4 Error Messages when any error message appears during the procedure.
5-3 LINEARITY CALIBRATION
.
.
....
Five point linearity calibration.
Choosing from 1P ~ 5P by
pressing key.
No linearity
calibration value at this point
linearity calibration
value present at this point
The weight value is displayed.
Increase the number of
the flashing digit by one
Decrease the number
of the flashing digit by
one
Shift the flashing digit
one space to left
Shift the flashing digit
one space to right
Save the settings
Exit the settings
Press and key to
choose linearity calibration
C A L
ZERO
+
TARE
ENTER
ZERO
+
TARE
ENTER
STANDBY
OPERAE
ENTER
STANDBY
OPERAE
ENTER
STANDBY
OPERAE
STANDBY
OPERAE
TARE
PRINT
ZERO
+
GROSS
NET
ENTER
STANDBY
OPERAE