MNL004_EN rev. F
Pag. 26 di 56
4.3 DMLP DIGITAL MONITOR OPERATION
The equipment can be supplied with DMLP Digital Monitor, completely independent from the ECT-F controller, to
record the temperatures, temperature alarms and power failures (thanks to a backup battery supplied as standard,
automatically re-charged.
When the DMLP Digital Monitor is switched on (see par. 5.2.1 point 5), it monitors and records every minute of
operation. Data is stored in two internal memories. The first memory is represented by DMLP Printer (if installed): it
grants a quick access to the last recorded 45 days. The second memory is an SD Secure Digital car, a black box that
grants at least five year of data storage.
The information stored on the SD Card is:
Day of the week, date and hour
Temperature chamber A and B (probe chamber A and probe chamber B)
Door status chamber A and B (probe chamber A and probe chamber B)
Door openings time chamber A and B (probe chamber A and probe chamber B)
Relay status (dry contact)
Battery voltage, mains presence/absence
High and low temperature limits and alarm delay time
Alarms and failures codes
Modified parameter code
DMLP Digital Monitor includes an internal lock to register all the events and no. 2 dry contacts + RS485 IN/OUT for
bi-directional remote control/monitoring (Web Light Server and/or GSM Communicator).